RRC ID 37030
著者 Nozoe T, Goda S, Selyanchyn R, Wang T, Nakazawa K, Hirano T, Matsui H, Lee SW.
タイトル In vitro detection of small molecule metabolites excreted from cancer cells using a Tenax TA thin-film microextraction device.
ジャーナル J Chromatogr B Analyt Technol Biomed Life Sci
Abstract We developed a new device for the in vitro extraction of small molecule metabolites excreted from cancer cells. The extraction device, which was biocompatible and incubated with cancer cells, consists of a thin Tenax TA film deposited on the surface of a cylindrical aluminum rod. The Tenax TA solid phase was utilized for the direct extraction and preconcentration of the small molecule metabolites from a cell culture sample. The device fabrication and the metabolite extraction were optimized, tested, and validated using HeLa cell cultures. Comparison of metabolic profiles with the control measurement from the culture medium enabled detection of metabolites that were consumed or produced by the cell culture. Tentative identification and semi-quantitative investigation of the excreted metabolites were performed by GC-MS analysis. The proposed approach can be a valuable tool for the characterization of low-volatile cancer cell metabolites that are not covered by use of conventional methods based on headspace solid phase microextraction.
巻・号 991
ページ 99-107
公開日 2015-6-1
DOI 10.1016/j.jchromb.2015.04.016
PII S1570-0232(15)00219-6
PMID 25932789
MeSH Cell Culture Techniques Equipment Design Gas Chromatography-Mass Spectrometry / methods HeLa Cells Humans Neoplasms / chemistry* Polymers Solid Phase Microextraction / instrumentation* Solid Phase Microextraction / methods Volatile Organic Compounds / analysis Volatile Organic Compounds / isolation & purification*
IF 3.004
引用数 13
WOS 分野 BIOCHEMICAL RESEARCH METHODS CHEMISTRY, ANALYTICAL
リソース情報
ヒト・動物細胞 HeLa(RCB0007)