RRC ID 38801
著者 Ohashi E, Karube I.
タイトル Development of a thin membrane glucose sensor using beta-type crystalline chitin for implantable biosensor.
ジャーナル J Biotechnol
Abstract A biocompatible thin membrane glucose sensor has been developed using beta-type crystalline chitin as a supporting material for the purpose of implantable biosensor. Gold electrode was formed by vapor deposition of the surface of glucose oxidase-immobilized beta-chitin membrane. The sensor showed rapid responses to glucose and H2O2 (< 45 s in batch operation), and addition of 2,6-dichlorophenolindophenol as a mediator caused expansion of the measuring range. To investigate the possible uses for the implantable, the sensor sterilization and toxicity were studied. Sterilization with ethylene oxide was achieved by exposing samples, and a decrease in activity was permissible for measurements of body fluid. From the results of the toxicity assay using mouse fibroblast cells, beta-chitin sensor, constructed with chitin, glucose oxidase and gold electrode, showed no particular toxicity.
巻・号 40(1)
ページ 13-9
公開日 1995-5-15
DOI 10.1016/0168-1656(95)00028-o
PII 016816569500028O
PMID 7794598
MeSH Animals Biosensing Techniques* Biotechnology Carbohydrate Conformation Cell Line Chitin* / chemistry Crystallization Enzymes, Immobilized Ethylene Oxide Evaluation Studies as Topic Glucose / analysis* Glucose Oxidase Gold Membranes, Artificial Mice Prostheses and Implants Sterilization
IF 3.503
引用数 27
WOS 分野 BIOTECHNOLOGY & APPLIED MICROBIOLOGY
リソース情報
ヒト・動物細胞 L929(RCB0081)