RRC ID 63828
著者 Tottori N, Nisisako T.
タイトル Particle/cell separation using sheath-free deterministic lateral displacement arrays with inertially focused single straight input.
ジャーナル Lab Chip
Abstract This paper proposes microfluidic particle separation by sheath-free deterministic lateral displacement (DLD) with inertial focusing in a single straight input channel. Unlike conventional DLD devices for size-based particle separation, in which sheath streams are used to focus the particles before the solution containing them reaches the DLD arrays, the proposed method uses inertial focusing to align the particles along the middle or the sidewalls of the straight rectangular input channel. The two-stage model of inertial focusing is applied to reduce the length of the side-focusing channel. The proposed method is demonstrated by using it to separate fluorescent polymer particles of diameters 13 and 7 μm, in the process of which the effect of the particle focusing regime on the separation performance is also investigated. Through middle focusing, the method is further used to separate MCF-7 cells (a model of circulating tumor cells (CTCs)) and blood cells, with ∼99.0% capture efficiency achieved.
巻・号 20(11)
ページ 1999-2008
公開日 2020-6-2
DOI 10.1039/d0lc00354a
PMID 32373868
MeSH Cell Separation Humans Microfluidic Analytical Techniques* Microfluidics Neoplastic Cells, Circulating* Particle Size
IF 6.774
リソース情報
ヒト・動物細胞 MCF7(RCB1904)